Using method:
1. Choose the appropriate size of the clip and the compatible applier.
2. Check the compatibility of all devices prior to use.
3. Following aseptic rules remove clips cartridge from the single packing. To prevent any damage of the device
place it on a sterile surface.
4. Grip the applier around the bolt (alike pencil is griped). For endo appliers grip applier around the shaft.
5. Align applier jaws vertically and laterally over a clip in the cartridge and advance product jaws into the slot of
the clip cartridge making sure they are perpendicular to the surface of the cartridge. Advance the jaws till they
are stopped. Do not use force to push the applier. The applier should move inside and outside of the slot easily.
6. Remove the applier from the cartridge. The clip is affixed in the jaws. It is not required to take any actions to
keep the clip in place.
7. Verify that the clip is fully inserted into the applier jaws and the clip legs do not protrude beyond the end of the
jaws.
8. Handle the applier carefully. The jaws should not close prematurely.
9. Place the clip around the structure intended for ligation or marking. Use appropriate force to close the clip
completely making sure it is placed properly. Closure should be made with a smooth, firm continuous motion
until clip is fully closed. Releasing the pressure on the handles will cause the applier jaws to spring open.
10. Remove the applier from the surgical site.
Size |
Detachable |
Length |
Diameter |
Ref |
M |
2 parts |
33cm |
10mm |
TATL2200E2 |
3 parts |
45cm |
10mm |
TATL2200E3 |
ML |
2 parts |
33cm |
10mm |
TATL3200E2 |
3 parts |
45cm |
10mm |
TATL3200E3 |
L |
2 parts |
33cm |
10mm |
TATL4200E2 |
3 parts |
45cm |
10mm |
TATL4200E3 |
Product display
Company Profile
Certifications
FAQ
1, Does the clip match Weck or Horizon brand?
Yes.
2Does your factory produce applier?
Yes, we produce appliers for open surgery and lap surgery.
3, How many sizes does Tiga clip have?
There are 6 sizes, Micro, S, MS, M, ML and L.
4, Does it have CE certificate?
Yes, at present the CE is under MDD, and we are making CE under MDR now.
https://conston.en.made-in-china.com/contact-info.html